Using the QT PC tool

More
12 years 11 months ago #725 by fusername
So I am using the QT PC software on ubuntu, worked right out of the box so that was a huge reliefe after seeing how much trouble WIN users must go thru to get it working.? However I am noticing an "issue" or maybe feature?? When I increase the gain on channel1, eventualy what was a readable voltage before is clipped flat.? I don't konw if this is in the pc software or as I suspect, the firmware of the scope, meaning at higher gain I can not read higher voltages without using AC coupling.
Also more importantly, the QT tool gives very incorrect voltage readings for the analog channels.? The scope seems right on its own, but the readings on screen are way off and change w/ the gain settings.? I tried to look at the source code but QT always confsues the bejeezus out of me, this stuff is way? beyond my ability or I'd try and help!

Please Log in or Create an account to join the conversation.

More
12 years 11 months ago - 12 years 11 months ago #726 by ganzziani
Replied by ganzziani on topic Re:Using the QT PC tool
Hi! The QT app is still in development, I'll check if there is a problem with the voltage readings. Did you verify the voltages on Xminilab's screen?
If you increase the gain a lot, then the waveform will look clipped.
The QT app is being updated regularly, hopefully next week I will make the first binary release.
Last edit: 12 years 11 months ago by ganzziani.

Please Log in or Create an account to join the conversation.

More
12 years 11 months ago #731 by fusername
Replied by fusername on topic Re:Using the QT PC tool
So the clipping occurs for what reason? Does the uC select a different ref voltage at higher gain levels? it doens't 'look' clipped, ie goes flat at the top of the screen, but 'is' clipped meaning if I move it down in the window it still plateaus where I know it is climbing higher. annoying since if it is b ecasue vref is made too low to measure I can't look at low freq signals with DC bias at a high gain level, but my leader scope has the same feature, so I am not too surpised to hear it is supposed to work that way, but I was hoping it was due to the analog nature of my scope, and taht this lil guy could do it./\r\n//\r\n/As for the voltage readings, they are deffinatly wrong, although I have not pulled latest and tried it today. What happens is if I put a horizontal cursor in to measure the voltage, the value it reads changes as I change the gain, even if the cursor is kept on top of the wave form. It can go as far as to claim 20v on what I know is a .18v signal. I'll test it with latest right now however. Would you prefere I submit bugs thru github? or would you prefer I just let you develop it on your own and wait for a real release?

Please Log in or Create an account to join the conversation.

More
12 years 11 months ago #732 by ganzziani
Replied by ganzziani on topic Re:Using the QT PC tool
So the clipping occurs for what reason?
The clipping occurs because there is too much gain./\r\n/There is only one reference voltage of 2.048V. Moving the waveform position is only done visually, there is no offset added to the signal.
As you know, to measure small signals on top of a DC bias, you would need to select AC coupling.
Are you reading wrong voltages from the PC app, or from the device's screen?
The voltage values read with a horizontal cursor will certainly change when changing the channel gain.
You can submit bugs here on the forum.

Please Log in or Create an account to join the conversation.

More
12 years 11 months ago #735 by fusername
Replied by fusername on topic Re:Using the QT PC tool
oh you know what, I was having a brain fart there. I was thinking the gain was done in software, I had forgotten the ATMEGA had actual internal gain circuitry.? as for the change in value, I think you can re-create what I am describing in a few simple steps.
?
Attach the scope to a stable voltage, in my case a disconnected sensor sitting at .182v (voltage read from physical scope in DMM mode)
plot this on the pc with gain slider in 4tth position from bottom
turn on horizontal curosr for the channel in question, align the cursor with the voltage, see VA reading of whatever (-.37 volts)
decrease? the gain slider on the pc one step
if the horizntaal curosr has shifted off the trace, re-align it and read the new voltage, it does not seem to move ever however
new Va reading is -.76
?
not sure why it is negative either.? also by the logic of my cursor, when the setting is 1.28 v/div, the horizontal cursor actually indicates 2.3v/div when I drag a cursor from one div to the next.? HOWEVER setting one cursor on one div and the other cursor on the next div, I get a reading like
Va -3v
Vb -5.3V
delta 1.3v
so the delta is right but the readouts are wrong....? I screencaped it for you.
?
I am in ubuntu 12.04

Please Log in or Create an account to join the conversation.

More
12 years 9 months ago #782 by fusername
Replied by fusername on topic Re:Using the QT PC tool
this bug appears resolved, thank you dev team!

Please Log in or Create an account to join the conversation.

Time to create page: 0.212 seconds
Powered by Kunena Forum