15 years 6 months ago - 11 years 11 months ago #139 by ganzziani
2.40:??? - Fixed bug with elastic mode/\r\n/2.39:??? - Doubled visible data in sniffer when using ASCII/\r\n/ - Valid fonts for Line Feed and Carriage return characters/\r\n/2.38??? - No change/\r\n/2.37??? - Fixed power down bugs from revision 2.28/\r\n/2.36??? - Added a timer on the Pulse Counter/\r\n/2.35??? - Implemented the Pulse Counter/\r\n/2.34??? - Removed exponential average from volt meter measurements to improve response time/\r\n/ - Fixed bugs with serial transmission/\r\n/2.33??? - Cursors can go beyond the visible screen/\r\n/2.32??? - Distinct Start of Frame for fast and slow sampling rates/\r\n/2.31??? - Added Start of Frame to the serial interface/\r\n/2.30??? - Added the METER measurement to the interface protocol/\r\n/2.29??? - Updated serial interface protocol/\r\n/2.28??? - Reference waveforms on XY mode/\r\n/2.27??? - No change2.26??? - Option to invert the Chip Select pin in the SPI sniffer/\r\n/2.25??? - Minor cleanup to the source code/\r\n/2.24??? - Added the "Restore Settings" command to the inteface protocol/\r\n/ - On the first power up, the device will enter the calibration function/\r\n/2.23??? - no change/\r\n/2.22??? - Optimization on I2C sniffer/\r\n/2.21??? - The Frequency Counter now works on any digital pin (selected with the trigger source)/\r\n/ - On the external trigger, the maximum frequency is 12MHz/\r\n/ - On the logic channel pins, the maximum frequency is 16MHz/\r\n/ - The trigger mark is shown when a logic input is the trigger source/\r\n/ - The trigger direction is toggled when selecting a digital trigger source/\r\n/ - Fixed SPI sniffer bug caused when data is being transmitted before the sniffer has started/\r\n/ - Added graphic indication of the SPI configuration in the menu/\r\n/ - Optimizations on UART sniffer/\r\n/2.20??? - Fixed bug created in 2.19 that disabled low level interrupts/\r\n/2.19??? - Fixed SPI Sniffer bugs:/\r\n/ - Fixed bug when receiving a very long stream of data/\r\n/ - Fixed bug when CPOL was 1/\r\n/ - Fixed bug that enabled the ASCII mode when CPHA was 1/\r\n/ - Verified Sniffer SPI clock operation range:/\r\n/ - Correct operation up to 439kHz/\r\n/ - Above 439kHz, MOSI still works up to 8MHz/\r\n/2.18??? - Overflow when displaying TPost has been fixed/\r\n/ - Fixed display flicker during sniffer mode/\r\n/2.17??? - UART Sniffer can decode 5,6,7 data bit sizes, in addition to 8 which was fixed./\r\n/ - Menu timeout after 1 minute/\r\n/2.16??? - Fixed bug on the UART sniffer/\r\n/2.15??? - Minor bug fixed in the XY mode when the ROLL is enabled./\r\n/2.14??? - Minor improvements to UI/\r\n/2.13??? - Prevent going to sleep when connected to USB/\r\n/2.12??? - Updated interface protocol. The PC interface needs to be updated as well./\r\n/2.11??? - Better handling when stopping the scope/\r\n/ - Added Trigger Timeout control/\r\n/2.10??? - USB restart when USB communication lost/\r\n/2.09??? - Code optimizations/\r\n/2.08??? - XY mode: 1.- The graph can be moved vertically, 2.- Cursors added/\r\n/2.07??? - If VCC is lower than about 3.15V: 1.- An empty battery icon will appear 2.- No writes to EEPROM/\r\n/2.06??? - Fixed bug in Frequency Counter/\r\n/2.05??? - Fxed bug in I2C sniffer/\r\n/2.04??? - No change/\r\n/2.03??? - Fixed display's power off sequence/\r\n/2.02??? - Extended reset signal when initializing the display, added delay before initializing display/\r\n/2.01??? - Extended reset signal when initializing the display/\r\n/2.00??? - Getting ready to send out a newsletter, reving up to 2.00
Older revision history is in the readme.txt file, along with the HEX file
Last edit: 11 years 11 months ago by ganzziani.

Please Log in or Create an account to join the conversation.

More
14 years 2 months ago #360 by Joseph Levy
Hi
Updated from ver 1.86  (HW 2.2)
Counter of Ext Trig counts only low frequencies ( 600-700 Hz) and counts way too high on higher freq. , like 2-4 times of real freq.  (feeding it with squre wave 4 v p-p fron 0 volts to 4 volts).
 
 TIA Joe.

Please Log in or Create an account to join the conversation.

More
14 years 2 months ago - 14 years 2 months ago #361 by ganzziani
I assume you are using the embedded waveform generator. Remember that by default the wave is centered at 0V, the external trigger pin expects digital signals. Try adding 1V offset to your wave, and set the amplitude at 2V. (The AWG range is +/- 2V). Since the AWG has a low pass filter, it will not work very well at high frequencies. My testing shows that if using the AWG, it works ok up to 95kHz. If you want to check out the frequency counter, you could use a 555 timer, or the output of a microcontroller.../\r\n/
Last edit: 14 years 2 months ago by ganzziani.

Please Log in or Create an account to join the conversation.

More
14 years 2 months ago #365 by Joseph Levy
Hello Gabriel
The test signal comes from a HP function generator and is 0 volts to 4 volt and the frequency shows correct to 255 Hz and after that jumps to much higher readings (although on a seperate oscilloscope and a counter it shows fine, both levels and frequency)
            

Please Log in or Create an account to join the conversation.

More
14 years 2 months ago #366 by ganzziani
Hmmm, it works flawlessly on my desk. I am using a BK Precision function generator, and I see the same frequency displayed on my Xminilab +/- 1Hz. It works great up to almost 13MHz.
Let's try to see what's wrong. Let's pick a frequency where it doesn't work, say 1kHz./\r\n/Check if the signal looks ok on the analog channels:
  1. Check with the Scope Mode that the Xminilab is indeed receiving 0V to 4V.
  2. Check with the FFT Mode at the 8kHz max setting, that you have perfect peaks at 1KHz, 3kHz, 5Khz, 7kHz.
  3. Check with the Meter mode the VDC, VPP, and Frequency reading the Meter, and confirm that they are 2V, 4V and 1kHz respectively.
Let me know...

Please Log in or Create an account to join the conversation.

More
14 years 2 months ago #367 by Joseph Levy
Hi
The culprit found! It's the BlueTooth dongle..
Once it was turned off the counter works fine.
Now I'll try to shield some wires and move the BT antenna side away from the inputs.

Please Log in or Create an account to join the conversation.